Military standard blowing dust blowing sand testing chamber | ||||||
Model | WBE-JPCSS1000L | |||||
Valume(L) | 1000L | |||||
Performance index | Test range | Landscape blowing dust and sand | ||||
Flow rate | 1.5~29m/s(Adjustable type) | |||||
Blow dust | Variety | concentrationg/m³ | Temperature℃ | Humidity% | Air speed m/s | |
Talc powder MIL STD DUST |
10.6±7g/m³ | 23~55℃ | ≦30%RH | 1.5~8.9m/s | ||
Blow sand | MIL STD SAND | 0.18+0.2g/m³ 1.1±0.3 g/m³ 2.2±0.5 g/m³ |
55~85℃ | 18~29 m/s | ||
Chamber material | Inner | SUS#304 stainless steel plate | ||||
Outer | SECC steel plate +2660 powder paint treatment | |||||
Control system | Fuji touch screen + Mitsubishi PLC program controller | |||||
Inner size (mm) |
(W) | 1600 | ||||
(D) | 800 | |||||
(H) | 800 | |||||
Housing size (mm) |
(W) | 7200 | ||||
(D) | 2700 | |||||
(H) | 2680 | |||||
Observation window | Double vacuum glass large observation window | |||||
Refrigerating system | 1. Imported high-efficiency refrigeration compressor 2. The whole system uses R404 environmental refrigerant 3. The whole system pipeline is tested for 48H by nitrogen pressure |
|||||
Power supply | AC(380±10)V 50Hz Three phase | |||||
Apply environment | 1.Temperature:5℃~30℃;2.Relative humidity:≤85%RH,Air pressure:86kPa~106kPa | |||||
Protect device | 1.Protection device for the door open;2.Blower overload protection;3.Compressor overload protection;4.Indicator protection | |||||
Meet implementation criteria | GJB150.12 / GJB11621.7ARepeated cycle dust blowing test specification | |||||
Instruction |